What does this quality measure capture?
A matrix spike adds a known amount of analyte to a sample aliquot before preparation and analysis. Comparing its result with the matching unspiked sample helps assess how the sample matrix affected recovery under the laboratory method.
The subtraction removes the concentration already present in the unspiked sample, leaving the recovered spike amount. The result is then compared with the amount added. This is a recovery calculation, not a correction to the reported sample concentration.
Define the cohort and units
Enter spiked result, unspiked result, and known amount added. All three concentrations must be nonnegative, on the same sample basis, and in compatible units; the amount added must be positive.
The calculator outputs percent recovery and the recovered spike amount in the concentration unit. A recovery above 100% is possible when the measured increment exceeds the amount added; a negative value can occur if the spiked result is below the unspiked result.
Formula and worked example
Recovery is 100×(SSR−SR)/SA, where SSR is the spiked-sample result, SR is the unspiked result, and SA is spike added. The calculator does not apply dilution factors or convert between sample mass, volume, and concentration bases.
If SSR=8.4 mg/L, SR=2.0 mg/L, and SA=6.0 mg/L, recovered spike is 6.4 mg/L and recovery is 6.4/6×100=106.67%. The result exceeds 100% without being automatically clipped.
Interpret the displayed result
If SSR=13.5 mg/L, SR=4.5 mg/L, and SA=10 mg/L, recovered spike is 9 mg/L and recovery is 90.00%. This value is arithmetic only; acceptance depends on the analyte method, concentration range, and laboratory QA plan.
The output rounds recovery to two decimals and recovered concentration to four. Keep original measurement precision and identify any dilution or blank correction already applied to the inputs so the calculation basis can be reproduced.
Assumptions and limits
EPA describes matrix spike recovery as one way to assess accuracy and matrix-specific bias. It is not a universal acceptance criterion; method-specific limits and the laboratory’s quality plan determine whether the value is acceptable for that analysis.
A low or high recovery may relate to matrix interference, extraction, digestion, instrument response, or sample preparation. The single ratio does not identify which mechanism occurred and should be considered with laboratory control samples and method records.
Keep the result reproducible
Use matched aliquots and follow the method’s spike amount, timing, mixing, and reporting basis. If a sample result is reported after dilution correction while the spike amount is not, the ratio will be inconsistent even if every numeric entry is positive.
Confirm that the unspiked and spiked results have matching blank treatment and dilution correction before subtraction. A mismatch in preprocessing can dominate the small increment being measured and make a plausible-looking percentage misleading.
Record the analyte, method, original and spiked aliquots, spike level, dilution factors, and method acceptance range. the matrix spike recovery calculator checks the ratio, while QA review determines data usability.
Worked examples
The calculations subtract the matching unspiked result before comparing the recovered increment with spike added. Keep the three values on one concentration and dilution basis.
Recovery slightly above one hundred percent
Inputs: SSR=8.4 mg/L; SR=2.0 mg/L; SA=6.0 mg/L
Calculation: Recovered amount=8.4−2.0=6.4 mg/L; recovery=6.4/6×100=106.6667%
The calculator displays 106.67% recovery and 6.4000 mg/L recovered spike. Do not clip the result to 100%.
Ninety percent recovery
Inputs: SSR=13.5 mg/L; SR=4.5 mg/L; SA=10 mg/L
Calculation: Recovered amount=13.5−4.5=9 mg/L; recovery=9/10×100=90%
The result is 90.00% recovery with 9.0000 mg/L recovered. Compare it with the method-specific acceptance range.