Analytical Detection and Quantitation Limit Calculator

Estimate analytical detection and quantitation limits from response variability and calibration slope.

4,000 tools
Analytical Detection and Quantitation Limit Calculator — Response standard deviation σ (response units), Calibration slope S (response units per concentration unit) and calculation method
Analytical Detection and Quantitation Limit Calculator: inputs and calculation method.

How to use this calculator

Estimate analytical detection and quantitation limits from response variability and calibration slope.

  1. Enter Response standard deviation σ (response units), Calibration slope S (response units per concentration unit). Match the units shown beside each field and review the result together with the assumptions below.
  2. Read the worked example below to check the expected input format and result.
  3. Compare another scenario by changing one input at a time. Use Reset to return to the starting values.
Chemistry calculation

How this calculation works

ICH Q2(R2) describes estimating analytical detection and quantitation limits from response standard deviation and calibration slope.

Inputs and units

responseSd — Response standard deviation σ (response units)
Standard deviation of the analytical response, estimated using the chosen validation approach.
slope — Calibration slope S (response units per concentration unit)
Positive slope from the analyte calibration curve.

Formula

Displayed formula

DL = 3.3 σ / S; QL = 10 σ / S.

Method

The calculator divides the response standard deviation σ by the positive calibration slope S and applies the ICH factors 3.3 and 10. Because S is response per concentration, both limits are reported in concentration units.

Worked example

With σ = 0.012 response units and S = 0.80 response units per mg/L, the estimated DL is 0.0495 mg/L and QL is 0.150 mg/L.

Assumptions and limits

This is the ICH response-standard-deviation/slope estimate for a linear calibration. It does not validate a method or replace confirmation of accuracy, precision, matrix effects, and the reportable range. Enter σ and S on compatible response and concentration scales.

Frequently asked questions

What does σ represent?

It is a response standard deviation estimated from an appropriate blank, calibration, or validation data set.

Why are the outputs in concentration units?

The response standard deviation is divided by a calibration slope measured in response per concentration.

Are these limits sufficient for method validation?

No. The guideline describes estimates; the analytical procedure still needs validation appropriate to its intended use.

How do I use the Analytical Detection and Quantitation Limit Calculator?

Enter Response standard deviation σ (response units), Calibration slope S (response units per concentration unit). Match the units shown beside each field and review the result together with the assumptions below.

What is a worked example of Analytical Detection and Quantitation Limit?

With σ = 0.012 response units and S = 0.80 response units per mg/L, the estimated DL is 0.0495 mg/L and QL is 0.150 mg/L.