Sauerbrey Quartz Crystal Microbalance Mass Calculator

Estimate rigid-film mass loading from a quartz crystal microbalance frequency shift using the Sauerbrey relation.

4,000 tools
Sauerbrey Quartz Crystal Microbalance Mass Calculator — Fundamental crystal frequency f0 (MHz), Measured frequency shift Δfn (Hz), Overtone number n and calculation method
Sauerbrey Quartz Crystal Microbalance Mass Calculator: inputs and calculation method.

How to use this calculator

Estimate rigid-film mass loading from a quartz crystal microbalance frequency shift using the Sauerbrey relation.

  1. Enter Fundamental crystal frequency f0 (MHz), Measured frequency shift Δfn (Hz), Overtone number n, Active deposit area (cm²). Match the units shown beside each field and review the result together with the assumptions below.
  2. Read the worked example below to check the expected input format and result.
  3. Compare another scenario by changing one input at a time. Use Reset to return to the starting values.
Physics calculation

How this calculation works

The Sauerbrey relation links a small, rigid surface mass change to the normalized frequency shift of a thickness-shear quartz resonator. A negative frequency shift corresponds to added mass.

Inputs and units

fundamentalMHz — Fundamental crystal frequency f0 (MHz)
Unloaded fundamental resonance frequency in MHz.
shiftHz — Measured frequency shift Δfn (Hz)
Signed measured shift at the entered overtone; mass loading usually gives a negative shift.
overtone — Overtone number n
Integer harmonic number for the measured frequency shift.
area — Active deposit area (cm²)
Area over which the rigid film is assumed to be uniformly distributed.

Formula

Displayed formula

Δm/A = −Cf(Δfn/n); Cf = 17.7(5 MHz/f0)² ng·cm⁻²·Hz⁻¹ for an AT-cut quartz crystal.

Method

The frequency change is divided by the entered overtone number before applying the 5 MHz mass-sensitivity constant scaled by the square of the fundamental-frequency ratio. Multiplying areal loading by active area gives the signed total mass change.

Worked example

For a 5 MHz crystal with a −5.7 Hz fundamental shift over 1 cm², the estimated loading is 100.89 ng/cm², or 0.10089 µg total.

Assumptions and limits

Use for a small, rigid, uniformly coupled film in air or vacuum. The model does not correct for liquid viscosity, viscoelastic films, large loading, electrode geometry, or nonuniform deposits. A positive frequency shift produces a negative inferred mass change.

Frequently asked questions

Why does a negative shift mean positive mass?

Added rigid mass lowers the resonance frequency, so the Sauerbrey equation includes a minus sign.

Why divide by the overtone?

For an ideal thin rigid film, the frequency shift scales with harmonic number. Normalizing by n lets the fundamental sensitivity constant be used.

Can I use this with a liquid sample?

Not reliably when viscous or viscoelastic loading contributes to the shift. Use a model that accounts for dissipation and liquid loading.

How do I use the Sauerbrey Quartz Crystal Microbalance Mass Calculator?

Enter Fundamental crystal frequency f0 (MHz), Measured frequency shift Δfn (Hz), Overtone number n, Active deposit area (cm²). Match the units shown beside each field and review the result together with the assumptions below.

What is a worked example of Sauerbrey Quartz Crystal Microbalance Mass?

For a 5 MHz crystal with a −5.7 Hz fundamental shift over 1 cm², the estimated loading is 100.89 ng/cm², or 0.10089 µg total.