How to use this calculator
Estimate rigid-film mass loading from a quartz crystal microbalance frequency shift using the Sauerbrey relation.
- Enter Fundamental crystal frequency f0 (MHz), Measured frequency shift Δfn (Hz), Overtone number n, Active deposit area (cm²). Match the units shown beside each field and review the result together with the assumptions below.
- Read the worked example below to check the expected input format and result.
- Compare another scenario by changing one input at a time. Use Reset to return to the starting values.
How this calculation works
The Sauerbrey relation links a small, rigid surface mass change to the normalized frequency shift of a thickness-shear quartz resonator. A negative frequency shift corresponds to added mass.
Inputs and units
fundamentalMHz— Fundamental crystal frequency f0 (MHz)- Unloaded fundamental resonance frequency in MHz.
shiftHz— Measured frequency shift Δfn (Hz)- Signed measured shift at the entered overtone; mass loading usually gives a negative shift.
overtone— Overtone number n- Integer harmonic number for the measured frequency shift.
area— Active deposit area (cm²)- Area over which the rigid film is assumed to be uniformly distributed.
Formula
Δm/A = −Cf(Δfn/n); Cf = 17.7(5 MHz/f0)² ng·cm⁻²·Hz⁻¹ for an AT-cut quartz crystal.
Method
The frequency change is divided by the entered overtone number before applying the 5 MHz mass-sensitivity constant scaled by the square of the fundamental-frequency ratio. Multiplying areal loading by active area gives the signed total mass change.
Worked example
For a 5 MHz crystal with a −5.7 Hz fundamental shift over 1 cm², the estimated loading is 100.89 ng/cm², or 0.10089 µg total.
Assumptions and limits
Use for a small, rigid, uniformly coupled film in air or vacuum. The model does not correct for liquid viscosity, viscoelastic films, large loading, electrode geometry, or nonuniform deposits. A positive frequency shift produces a negative inferred mass change.
Frequently asked questions
Why does a negative shift mean positive mass?
Added rigid mass lowers the resonance frequency, so the Sauerbrey equation includes a minus sign.
Why divide by the overtone?
For an ideal thin rigid film, the frequency shift scales with harmonic number. Normalizing by n lets the fundamental sensitivity constant be used.
Can I use this with a liquid sample?
Not reliably when viscous or viscoelastic loading contributes to the shift. Use a model that accounts for dissipation and liquid loading.
How do I use the Sauerbrey Quartz Crystal Microbalance Mass Calculator?
Enter Fundamental crystal frequency f0 (MHz), Measured frequency shift Δfn (Hz), Overtone number n, Active deposit area (cm²). Match the units shown beside each field and review the result together with the assumptions below.
What is a worked example of Sauerbrey Quartz Crystal Microbalance Mass?
For a 5 MHz crystal with a −5.7 Hz fundamental shift over 1 cm², the estimated loading is 100.89 ng/cm², or 0.10089 µg total.
Reference: Peer-reviewed practical Sauerbrey model for quartz crystal microbalances. See our calculation methodology.