What the two outputs mean
Each input should already be a standard uncertainty expressed on the same measurement scale. If components are independent, their variances add: square every component, add those squares, and take the square root. The result, u_c, has the same unit as the inputs. This is the root-sum-square method used to combine standard uncertainties; it is not the arithmetic sum of the component magnitudes.
The second result is U = k × u_c. It scales the combined standard uncertainty by the coverage factor k that you choose for the reporting purpose. A common convention is k = 2, but the calculator does not infer a confidence level from k. The meaning of an interval depends on the uncertainty model, distribution assumptions, degrees of freedom, and the requirements of the measurement task.
Prepare a compatible uncertainty budget
Enter two to fifty nonnegative components separated by commas. Convert every value to one unit before combining it: for example, 0.12 mm and 80 μm cannot be entered unchanged in the same list. Convert 80 μm to 0.08 mm first. The values are magnitudes of standard uncertainty, not signed corrections; a correction belongs in the measurement result, while its remaining uncertainty belongs in the budget.
The calculator treats the listed values as independent and applies no sensitivity coefficients or covariance terms. That is appropriate only when the components have already been expressed as contributions to the result and their correlation is negligible or otherwise handled. If a source affects several inputs, blindly combining them as independent can overstate or understate the combined uncertainty. Review the measurement model and covariance before using the number in a report.
Use the combined measurement uncertainty calculator after assembling a budget. Keep the component names, sources, conversion factors, evaluation method, units, and date with the calculation. The list alone is not an uncertainty budget: someone reviewing it should be able to understand where each component came from and why it is expressed as a standard uncertainty.
Worked examples
The examples use compatible units and retain enough digits to show the arithmetic. The displayed calculator output rounds to eight decimal places, while a report should use precision justified by the measurement process.
Three independent components in millimetres
Inputs: Standard components: 0.12 mm, 0.08 mm, 0.05 mm; k = 2
Calculation: u_c = √(0.12² + 0.08² + 0.05²) = √0.0233 = 0.15264338 mm; U = 2 × 0.15264338 = 0.30528675 mm
Combined standard uncertainty ≈ 0.15264338 mm; expanded uncertainty ≈ 0.30528675 mm at the entered k.
Components expressed in micrometres
Inputs: Standard components: 6 μm, 8 μm, 10 μm; k = 2.5
Calculation: u_c = √(6² + 8² + 10²) = √200 = 14.14213562 μm; U = 2.5 × 14.14213562 = 35.35533906 μm
Combined standard uncertainty ≈ 14.14213562 μm; expanded uncertainty ≈ 35.35533906 μm at the entered k.
How to interpret the result
Combined standard uncertainty describes the estimated spread associated with the result under the stated model. Expanded uncertainty creates a wider reporting interval by applying k. It does not correct a result, remove bias, or establish that the measurement process is fit for a particular tolerance. If two processes have the same u_c but different distributions or degrees of freedom, the same k may not represent the same coverage probability.
Compare U with a tolerance only after checking that they refer to the same measurand, unit, measurement conditions, and coverage basis. A small ratio may prompt a more capable method or an explicit conformity decision rule; this calculator does not choose either. Record the uncertainty statement together with the coverage factor and the method used to obtain the components.
Assumptions and limits
The calculator uses u_c = √Σuᵢ² and U = k u_c. It assumes a linear combination of already-scaled contributions with no entered covariance; the more general propagation law includes covariance terms and can use sensitivity coefficients. It also does not estimate a component from raw repeat readings, infer a distribution, calculate effective degrees of freedom, or decide whether a selected k is suitable.
Treat the results as a transparent arithmetic check, not as certification of a measurement. If the component list mixes standard, expanded, tolerance, or resolution values, convert or evaluate them correctly before entering them. If a procedure specifies a different model or reporting convention, follow that documented method and use the calculator only as a cross-check.