What repeatability and reproducibility describe
Repeatability is the variation associated with the equipment under the study conditions; reproducibility is the variation associated with appraisers or other reproducible conditions. The calculator takes their standard deviations as inputs. It does not estimate them from individual readings, so the values must already come from an appropriate study design and analysis.
The part-to-part standard deviation describes the observed spread among the parts in that study. It matters because the same measurement error can look small beside a wide range of parts and large beside a narrow range. A result therefore describes this study’s measurement system and sampled parts, not a permanent grade for the instrument in every use.
Follow the variation calculation
The combined repeatability-and-reproducibility standard deviation is GRR = √(EV² + AV²). Total study standard deviation is √(GRR² + PV²). The displayed study-variation percentage is 100 × GRR ÷ total study standard deviation. These are root-sum-square combinations, which assume the supplied components are comparable independent estimates on one measurement scale.
A separate tolerance comparison is 100 × 6 × GRR ÷ tolerance width. The factor of six expresses a six-standard-deviation spread under a normal model; it is not a confidence interval or measured defect rate. Distinct categories are floor(1.41 × PV ÷ GRR), a coarse indicator rounded down to a whole category.
Enter components on a common scale
Enter EV, AV, and PV as standard deviations in the same units and for the same characteristic. If one component is in micrometres and another is in millimetres, convert first. The tolerance is the positive specification width, USL − LSL, in those same units. The form accepts decimal variation values, but the analysis still depends on how the components were estimated.
Crossed, nested, destructive, short-term, and long-term studies can estimate different things. Do not combine incompatible components just because they share a unit. Keep the study report’s part selection, operator count, trial structure, and interaction treatment consistent.
Read the three comparison outputs
Study variation asks what share of the combined observed study spread is attributable to the measurement system under this simplified component model. Percent tolerance asks how the gage spread compares with the engineering tolerance. Neither percentage answers whether a measurement system is acceptable without the decision risk, product function, customer rules, and study method.
The NDC value is a floored estimate of distinct categories across the observed part spread. A small value means the measurement system may have difficulty separating nearby parts in this sample. It does not mean the gauge can literally label only that many unique dimensions, and it is not a substitute for reviewing the raw study, plots, and variance components.
Use the estimate with study context
A component-based calculation omits bias, linearity, stability over time, calibration traceability, resolution, operator-by-part interaction, and uncertainty in the input estimates unless those effects are already represented in the supplied components. A single standard deviation can also hide non-normal error, drift, or a subgroup pattern that matters operationally.
Treat the output as a transparent calculation from four supplied values. Review the measurement procedure, the range of parts, and the intended tolerance decision with the quality plan. If a study’s analysis used variance components, ANOVA, or a particular multiplier convention, keep that convention consistent rather than mix it with this simplified standard-deviation formula.
Worked examples
Each example uses one measurement unit; inputs are illustrative, not acceptance criteria.
Wider observed part spread
Inputs: EV = 0.18 mm; AV = 0.12 mm; PV = 0.90 mm; tolerance width = 6.00 mm.
Calculation: GRR = √(0.18² + 0.12²) = 0.21633 mm; total = √(0.21633² + 0.90²) = 0.92563 mm; study variation = 100 × 0.21633 ÷ 0.92563 = 23.37%; tolerance = 100 × 6 × 0.21633 ÷ 6.00 = 21.63%; NDC = floor(1.41 × 0.90 ÷ 0.21633) = 5.
The measurement-system spread is 23.37% of this study’s total variation and 21.63% of the stated tolerance under the calculator’s conventions.
A second system with a different tolerance
Inputs: EV = 0.10 mm; AV = 0.24 mm; PV = 0.80 mm; tolerance width = 5.50 mm.
Calculation: GRR = √(0.10² + 0.24²) = 0.26000 mm; total = √(0.26² + 0.80²) = 0.84119 mm; study variation = 100 × 0.26 ÷ 0.84119 = 30.91%; tolerance = 100 × 6 × 0.26 ÷ 5.50 = 28.36%; NDC = floor(1.41 × 0.80 ÷ 0.26) = 4.
The larger tolerance ratio and lower category count are consequences of these inputs; they are not a pass/fail decision by themselves.