Keep the values tied to one PT round
Use the participant result x and assigned value Xpt for the same measurand, proficiency-test item, unit, and round. The scheme supplies the standard deviation for proficiency assessment σpt and the assigned value’s standard uncertainty u(Xpt). The participant-result uncertainty u(x) should be a standard uncertainty for the reported result, not an expanded uncertainty copied without conversion.
These inputs come from different roles in the comparison. The assigned value and its uncertainty belong to the provider’s statistical model; the participant result and its uncertainty belong to the laboratory. Do not substitute a tolerance, standard error, or confidence interval unless it is converted to the exact quantity the score requires.
Understand the two denominators
The z-prime score is z′ = (x − Xpt) / √(σpt² + u(Xpt)²). Its denominator combines the performance criterion with uncertainty in the assigned value. The zeta score is ζ = (x − Xpt) / √(u(x)² + u(Xpt)²). Its denominator combines the two measurement uncertainties. Both retain the signed difference in the numerator.
Because the denominators are different, z-prime and zeta can differ substantially even for the same result. A score near zero means the result is close to the assigned value relative to that denominator; a large magnitude indicates a larger standardized difference, but the scheme’s own decision rules determine interpretation.
Check units and nonzero uncertainty
Enter participant result, assigned value, and all uncertainty terms in the same measurement unit. σpt and the standard uncertainties must be nonnegative. If every term in a denominator is zero, the corresponding score is undefined; use the provider’s defined assessment parameters instead of forcing a division.
Check whether the PT provider reports standard or expanded uncertainty and whether coverage factors are specified. Also verify that the assigned value is valid for the method or measurement range being assessed. The calculator checks the arithmetic inputs but cannot determine whether those quantities are appropriate for the scheme.
Interpret sign and magnitude in context
A positive score means x exceeds Xpt; a negative score means it falls below it. The absolute value measures the distance in units of the selected combined scale. A sign is not a direction-of-quality judgment unless higher or lower has a meaningful interpretation for the measurand.
Do not apply a generic pass/fail threshold across schemes. The PT provider defines the assessment criterion, and the relevant standard or scheme protocol explains how results are evaluated. A score is not a probability that the laboratory is wrong and is not a standalone estimate of measurement uncertainty.
Use the scores as part of investigation
When a result is flagged by the scheme, review the measurement process, raw observations, calibration, sample handling, and uncertainty budget. Compare z-prime and zeta carefully: disagreement can reflect a large or small uncertainty estimate, not necessarily a change in the measured result. Keep the actual inputs and formula with the report.
The calculator does not account for correlations between uncertainty components, apply a provider’s acceptance bands, or determine corrective action. Use the relevant PT instructions and quality system. For formal decisions, preserve the assigned-value certificate, uncertainty basis, result units, and applicable edition of the proficiency-testing guidance.
Worked examples
Both scores use the same signed result difference but different combined standard uncertainties.
Result above the assigned value
Inputs: x = 103.2 mg/L; Xpt = 100.0 mg/L; σpt = 2.0 mg/L; u(Xpt) = 1.0 mg/L; u(x) = 1.5 mg/L.
Calculation: Difference = 3.2 mg/L; z′ = 3.2 ÷ √(2.0² + 1.0²) = 1.431; ζ = 3.2 ÷ √(1.5² + 1.0²) = 1.775.
Both scores are positive because the participant value is above the assigned value. The larger ζ follows from its smaller uncertainty denominator in this example.
A smaller offset with different uncertainty
Inputs: x = 50.6 units; Xpt = 50.0 units; σpt = 1.0; u(Xpt) = 0.3; u(x) = 0.8.
Calculation: Difference = 0.6; z′ = 0.6 ÷ √(1.0² + 0.3²) = 0.575; ζ = 0.6 ÷ √(0.8² + 0.3²) = 0.702.
The scores remain below one in magnitude here, but scheme-specific criteria—not a universal threshold—govern any proficiency judgment.